0 part(s), total €0.00
Connection flange | 2" Tri Clamp |
---|---|
Resolution | 0.003 % F.S. |
Output signal: Measuring range | 4 – 20 mA |
Bakeout temperature at the flange | ≤ 200 °C |
Operating temperature | ≤ 160 °C |
Pressure max. | 2600 hPa |
Attribute | 2 switch points |
Input voltage(s) | 21 – 30 V DC |
Electrical connection | D-Sub, 15-pole, male |
Full scale | 1 hPa |
Weight | 897 g |
Power consumption max. | 18 W |
Materials in contact with media |
|
Measuring range | 1E-4 – 1E0 hPa |
Accuracy of measurement in range 1 | 0.4 % (of measured value) |
Measuring method |
|
Response time | 30 ms |
Switch-point | 0 – 99 % F. S. |
Number of switch-points | 2 pcs. |
Switch-point: Voltage DC | 30 V DC |
Switch-point: Current DC | 0.5 A |
Protection degree |
|
Temperature: Storage | -40 – 65 °C |
Temperature effect: on range | 0.02 % of measured value/ °C |
Temperature effect: on zero point | 0,005 % F.S./°C |
Ambient temperature | 10 – 50 °C |
Volume | 4.2 cm³ |